Triple Wavelength dual Color Potometer Model MoniSpec-C-AT
Description:
The triple Wavelength dual Color Potometer model MoniSpec-C-AT simultaneously detects two colors in liquids.
Configurable for the measurement of:
Lovibond red/yellow, APHA or Hazen color number, ASTM1500 / Saybolt, ppm, mg/l, EBC, …
Measuring principle Color Photometer:
The sensor records the absorption at three different wavelengths to detect the colors of liquids. Three detectors with different sensitivities and a gas-filled full-spectrum measuring lamp are installed in the sensor optics.
Measuring channel / detector 1: Detects the absorption caused by turbidity and the first color.
Measuring channel / detector 2: Detects the absorption caused by turbidity and the second color.
Reference channel / detector 3: Detects the turbidity only.
The difference between the detector signals ([color + turbidity] – turbidity) results in the respective color measurement values.
Triple Wavelength dual Color Potometer Model MoniSpec-C-AT
Features:
Triple wavelength absorption:
– The simultaneous measurement of absorbance at three different wavelengths.
– Enables precise measurements
– Compensation for the effects of turbidity, window contamination, lamp ageing, …
Calibration / maintenance interval typically 12 months (depending on application):
– Less effort for maintenance and longer service life.
– As a rule, calibration and maintenance are required every 12 months.
Material Measuring cell typically stainless steel 1.4404 (optionally titanium, PTFE, …):
– The measuring cell is made of stainless steel.
– Other materials such as titanium, Hastelloy, Peek or PTFE are optionally available.
Measuring window material:
– Sapphire is used as the window material.
– Other materials such as borosilicate or Plexiglas are optionally available.
Sealing material FKM:
– The seals are made of FKM.
– Optionally, seals made of other materials such as FFKM, EPDM or silicone are available.
Process connections: DIN flanges, ANSI flanges, NPT, …:
– Various connection options for different process requirements.
Measuring range span: 0 – 0.1 AU up to 0 – 5 AU:
– The measuring range extends from 0 to 0.1 absorption units up to 0 – 5 absorption units.
Suitable for use in Ex-Zone I & Ex-Zone II (ATEX-Ex d IIC / T6):
– With the optional Ex- Equippement, the device can be used safely in hazardous areas.
Optional cleaning Jets:
– A probe with nozzles for cleaning the measuring cell can be installed as an option.
Cleaning is CIP-capable:
– The device is suitable for automated cleaning in the pipe (CIP).