Inline Absorption Photometer / Turbidity Sensor Model MoniSpec-C-A

Description:

The process absorption photometer model MoniSpec-C-A measures the turbidity of liquids in process pipes.

Configurable for the measurement of:

EBC / TEF/ FTU /ppm /mg/l / g/l / % TS/ AU/ CU/ OD / …

Measuring principle:

The inline turbidity sensor model MoniSpec-C-A uses the principle of near-infrared light absorption (NIR) to detect turbidity in liquids. A beam of light shines through the product to be measured. A detector is positioned opposite of the light source and detects the NIR component emanating from the light source. The detected NIR radiation changes depending on the respective turbidity.

Low turbidity = high intensity
High turbidity = low intensity

The measuring transmitter evaluates the intensity signals and calculates the resulting measuring values.

Inline Turbidity Sensor Model MoniSec-C-A
Absorption Photometer for Turbidity- / Color- measuring Model MoniSpec-C-A
Turbidimetry Transmitter Model C1
Masuring Transmitter Model C1

NIR absorption measurement method:
– The measurement of absorption will be in the near infrared spectrum.
– IR absorption allows precise measurements and compensates for influences caused by changes in color.
– The sensor can be configured to measure at any wavelength in the visible spectrum (color measurement).
Low maintenance / long service life:
– Less maintenance and longer service life due to optimized measuring cell with Sapphire measuring windows.
– The measuring cell has been optimized for easy maintenance and cleaning.
Calibration / maintenance interval typically 12 months (depending on application):
– As a rule, the calibration should be checked every 12 months.
– However, the maintenance interval depends on the measuring range and application.
Measuring lamp, gas-filled (λ 380 – 1100nm), service life approx. 1 year:
– The measuring lamp is filled with gas and has a service life of approx. 1 year.
Suitable for use in Ex zone I & Ex zone II (ATEX-Ex d IIC / T6):
– With the optional Ex equipment, the device can be used safely in hazardous area.
Measuring cell material typically stainless steel 1.4404 / 316L
– The measuring cell is made of stainless steel.
Sealing material: FKM (optionally FFKM, EPDM, silicone, …):
– The seals are made of FKM.
– Seals made of other materials such as EPDM, FFKM or silicone are optionally available.
Process connections: DIN flanges, ANSI flanges, NPT, …:
– Various connection options for different process requirements.