Inline Turbidity Probe Model ProScat12

Description:

The rod probe model ProScat12 measures the turbidity of liquids in process. The probe is inserted into a pipeline either using a mounting flange or alternatively via a 2” ball valve with pipe radius and weld-on connection piece is used for installation.

Turbidity Measuring:

Turbidity is not a clearly defined quantity such as temperature, weight or pressure it is a subjective impression and therefore turbidity meters are typically calibrated with a reference standard such as formazine

Measuring Principle:

The probe uses the principle of 12° forward scattering to detect turbidity in liquids. Measuring principle forward scattered light. This method is used for turbidity meters providing a measuring range from approx. 0-1ppm up to 0-4000ppm. However, this specification can vary greatly depending on the application and the product to be measured. When measuring scattered light, a distinction is made between forward scattering, sideways scattering (nephelometry) and reflection.

Turbidity probe model ProScat12
Turbidity Probe Model ProScat12
Turbidimetry Transmitter Model C1
Transmitter for Turbidity measurements Model C1

Features of the measuring System:

Low maintenance / long service life:
– Compensation of eventual coatings at the measuring windows.
– Insensitive for Color changes of the measured product.
Calibration / maintenance interval typically 24 months (depending by application):
– As a rule, calibration is required every 24 months
– However, the maintenance interval depends on the measuring range and application.
NIR LED, service life approx. 5 years:
– The high power LED and typically has a service life of around 5 years.
– Alternatively, a gas-filled full spectrum measuring lamp is available with a service life of approx. 1-2 years.
Material typically stainless steel 1.4404
– Wetted parts of the rod probe is typically made of stainless steel.
– Other materials such as titanium, Hastelloy, Peek, carbon fiber-reinforced PTFE, PVDF, etc. are available as alternatives.
Sealing Material: 
– The seals are made of FKM.
– Optionally, seals made of other materials such as EPDM, FFKM or silicone are available on request.
Installation:
– In basin, open channel or tank using mounting clamps.
– Pipeline installation, porting via 2″ ball valve.
Configurable for measurement of: EBC / TEF/ FTU /ppm /mg/l / g/l / …
– The measuring system can be calibrated in any unit of measurement.